A procedure for preparing a Chromosorb 101 liquid chromatographic column is described. Some column parameters such as the permeability, efficiency and peak asymmetry factor were calculated. The elution of some benzene and phenol derivatives from the polymeric column with different mobile phases was performed. The results showed that some gas chromatographic adsorbents such as Chromosorb 101 can be used as packings for high-performance liquid chromatographic columns. The behaviour of these styrene copolymers resembles that of a C18 silica bonded phase. © 1987.
Chromosorb 101 as a packing material for reversed-phase chromatography
BLO, Gabriella;
1987
Abstract
A procedure for preparing a Chromosorb 101 liquid chromatographic column is described. Some column parameters such as the permeability, efficiency and peak asymmetry factor were calculated. The elution of some benzene and phenol derivatives from the polymeric column with different mobile phases was performed. The results showed that some gas chromatographic adsorbents such as Chromosorb 101 can be used as packings for high-performance liquid chromatographic columns. The behaviour of these styrene copolymers resembles that of a C18 silica bonded phase. © 1987.File in questo prodotto:
Non ci sono file associati a questo prodotto.
I documenti in SFERA sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.