CE with contactless conductivity detection has been used to determine the glycine neurotransmitter in periaqueductal gray matter (PAG) of rats. The LOD for glycine has been decreased to a value of 0.2 μM by adding 75% v/v of ACN to the samples and increasing the sample zone introduced to a value of 20% of the overall capillary length. The repeatabilities of the analyte migration times and the zone areas amount to 2.1 and 2.7%, respectively. The optimized CE/contactless conductivity detection method makes it possible to determine the micromolar concentrations of glycine in PAG microdialyzates without the necessity of sample derivatization. It follows from a pharmacological study that a local inflammation initiated by an application of carrageenan increased the glycine concentration in the rat PAG seven times, compared with a control. The glycine level in PAG can be decreased and the pain suppressed by administering paracetamol.
CE with contactless conductivity detection has been used to determine the glycine neurotransmitter in periaqueductal gray matter (PAG) of rats. The LOD for glycine has been decreased to a value of 0.2 μM by adding 75% v/v of ACN to the samples and increasing the sample zone introduced to a value of 20% of the overall capillary length. The repeatabilities of the analyte migration times and the zone areas amount to 2.1 and 2.7%, respectively. The optimized CE/contactless conductivity detection method makes it possible to determine the micromolar concentrations of glycine in PAG microdialyzates without the necessity of sample derivatization. It follows from a pharmacological study that a local inflammation initiated by an application of carrageenan increased the glycine concentration in the rat PAG seven times, compared with a control. The glycine level in PAG can be decreased and the pain suppressed by administering paracetamol. © 2009 Wiley-VCH Verlag GmbH & Co. KGaA.
A determination of submicromolar concentrations of glycine in periaqueductal gray matter microdialyzates using capillary zone electrophoresis with contactless conductivity detection
Soukupova MMembro del Collaboration Group
;
2009
Abstract
CE with contactless conductivity detection has been used to determine the glycine neurotransmitter in periaqueductal gray matter (PAG) of rats. The LOD for glycine has been decreased to a value of 0.2 μM by adding 75% v/v of ACN to the samples and increasing the sample zone introduced to a value of 20% of the overall capillary length. The repeatabilities of the analyte migration times and the zone areas amount to 2.1 and 2.7%, respectively. The optimized CE/contactless conductivity detection method makes it possible to determine the micromolar concentrations of glycine in PAG microdialyzates without the necessity of sample derivatization. It follows from a pharmacological study that a local inflammation initiated by an application of carrageenan increased the glycine concentration in the rat PAG seven times, compared with a control. The glycine level in PAG can be decreased and the pain suppressed by administering paracetamol. © 2009 Wiley-VCH Verlag GmbH & Co. KGaA.I documenti in SFERA sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.