We present a comparison between full field digital mammography and synthetic mammography, performed on several mammography systems from four different manufacturers. The analysis is carried out on both the digital and synthetic images of two commercially available mammography phantoms, and focuses on a set of objective metrics that encode the geometrical appearance of imaging features of diagnostic interest. In particular, we measured sizes and contrasts of several clusters of microcalcification specks, shapes and contrasts of circular masses, and the power spectrum of background regions mimicking the heterogeneous texture of the breast parenchyma. Despite the potential issues of tomosynthesis in terms of image blurring, the synthetic images do not highlight any globally significant differences in the rendering of the details of interest, when compared to the original digital mammograms: relative contrasts are generally preserved, as well as the geometry of broad structures. We conclude that, as far as the considered objective metrics are concerned, the image quality of synthetic mammography does not exhibit significant differences with respect to the one of full field digital mammography, for all the considered systems.

A comparative study of physical image quality in digital and synthetic mammography from commercially available mammography systems

CONTILLO, Adriano;Taibi, Angelo
Ultimo
2018

Abstract

We present a comparison between full field digital mammography and synthetic mammography, performed on several mammography systems from four different manufacturers. The analysis is carried out on both the digital and synthetic images of two commercially available mammography phantoms, and focuses on a set of objective metrics that encode the geometrical appearance of imaging features of diagnostic interest. In particular, we measured sizes and contrasts of several clusters of microcalcification specks, shapes and contrasts of circular masses, and the power spectrum of background regions mimicking the heterogeneous texture of the breast parenchyma. Despite the potential issues of tomosynthesis in terms of image blurring, the synthetic images do not highlight any globally significant differences in the rendering of the details of interest, when compared to the original digital mammograms: relative contrasts are generally preserved, as well as the geometry of broad structures. We conclude that, as far as the considered objective metrics are concerned, the image quality of synthetic mammography does not exhibit significant differences with respect to the one of full field digital mammography, for all the considered systems.
2018
Baldelli, Paola; Bertolini, Marco; Contillo, Adriano; Della Gala, Giuseppe; Golinelli, Paola; Pagan, Laura; Rivetti, Stefano; Taibi, Angelo
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in SFERA sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11392/2392493
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 13
  • ???jsp.display-item.citation.isi??? 12
social impact