Most conventional aerosol neutralizers are based on radioactive sources, which are controlled by strict regulations restricting their handling, transport, and storage. The TSI 3087 soft X-ray (SXR) neutralizer circumvents these legal restrictions. The aim of the present work is to compare the performance of a standalone SXR aerosol neutralizer with that of conventional radioactive aerosol neutralizers based on85Kr (TSI 3077) and241Am (Grimm 5522) by performing field tests in a real environmental scenario. The results obtained when the SXR neutralizer was connected to a mobility particle sizer spectrometer (MPS), different from the device suggested by the manufacturer, were comparable with those obtained with the use of radioactive aerosol neutralizers. In changing the neutralizer, the particle number concentrations, measured with the MPS connected to the SXR neutralizer, almost remained within the 10% uncertainty bounds for the particle size interval 10–300 nm, when diffusion losses inside the SXR tube were considered. Based on our comparisons, the SXR neutralizer can be regarded as a standalone instrument that could solve the problems associated with legal restrictions on radioactive neutralizers and fulfil the need for a portable instrument for different field test purposes.
Data di pubblicazione: | 2018 | |
Titolo: | Field study of a soft X-ray aerosol neutralizer combined with electrostatic classifiers for nanoparticle size distribution measurements | |
Autori: | Nicosia, Alessia; Manodori, Laura; Trentini, Arianna; Ricciardelli, Isabella; Bacco, Dimitri; Poluzzi, Vanes; Di Matteo, Lorenza; Belosi, Franco | |
Rivista: | PARTICUOLOGY | |
Keywords: | Aerosol neutralizer; Diffusion losses; Mobility particle sizer spectrometer; Particle size distribution; Soft X-ray non-radioactive charger; Chemical Engineering (all); Materials Science (all); Condensed Matter Physics | |
Abstract in inglese: | Most conventional aerosol neutralizers are based on radioactive sources, which are controlled by strict regulations restricting their handling, transport, and storage. The TSI 3087 soft X-ray (SXR) neutralizer circumvents these legal restrictions. The aim of the present work is to compare the performance of a standalone SXR aerosol neutralizer with that of conventional radioactive aerosol neutralizers based on85Kr (TSI 3077) and241Am (Grimm 5522) by performing field tests in a real environmental scenario. The results obtained when the SXR neutralizer was connected to a mobility particle sizer spectrometer (MPS), different from the device suggested by the manufacturer, were comparable with those obtained with the use of radioactive aerosol neutralizers. In changing the neutralizer, the particle number concentrations, measured with the MPS connected to the SXR neutralizer, almost remained within the 10% uncertainty bounds for the particle size interval 10–300 nm, when diffusion losses inside the SXR tube were considered. Based on our comparisons, the SXR neutralizer can be regarded as a standalone instrument that could solve the problems associated with legal restrictions on radioactive neutralizers and fulfil the need for a portable instrument for different field test purposes. | |
Digital Object Identifier (DOI): | 10.1016/j.partic.2017.08.001 | |
Handle: | http://hdl.handle.net/11392/2388277 | |
Appare nelle tipologie: | 03.1 Articolo su rivista |
File in questo prodotto:
File | Descrizione | Tipologia | Licenza | |
---|---|---|---|---|
Field study of a soft X-ray aerosol neutralizer combined with electrostatic classifiers for nanoparticle size distribution measurements.pdf | Full text editoriale | Full text (versione editoriale) | NON PUBBLICO - Accesso privato/ristretto | Administrator Richiedi una copia |