We have implemented a novel scheme of signal readout for resonant gravitational wave detectors. For the first time, a capacitive resonant transducer has been matched to the signal amplifier by means of a tuned high Q electrical resonator. The resulting 3-mode detection scheme widens significantly the bandwidth of the detector. We present here the results achieved by this signal readout equipped with a two-stage SQUID amplifier. Once installed on the AURIGA detector, the one-sided spectral sensitivity obtained with the detector operated at 4.5 K is better than 10-20Hz-1/2 over 110 Hz, and in good agreement with the expectations.
3-Mode Detection for Widening the Bandwidth of Resonant Gravi- tational Wave Detectors
F. Salemi;
2005
Abstract
We have implemented a novel scheme of signal readout for resonant gravitational wave detectors. For the first time, a capacitive resonant transducer has been matched to the signal amplifier by means of a tuned high Q electrical resonator. The resulting 3-mode detection scheme widens significantly the bandwidth of the detector. We present here the results achieved by this signal readout equipped with a two-stage SQUID amplifier. Once installed on the AURIGA detector, the one-sided spectral sensitivity obtained with the detector operated at 4.5 K is better than 10-20Hz-1/2 over 110 Hz, and in good agreement with the expectations.File in questo prodotto:
Non ci sono file associati a questo prodotto.
I documenti in SFERA sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.